WPC150 Portable Particle Counter is made in the principle of light extinction (or light blockage) and is used to detect the size and quantity of solid particles in liquid. It is widely used in aviation, space, electricity, oil, chemical, transportation, port, metallurgy, machinery and automobile manufacturing for detecting contamination of solid particles for hydraulic oil, lubricating oil, transformer oil, turbo generator oil (turbine oil), gear oil, engine oil, jet fuel and water based hydraulic oil and insoluble particles for organic liquid and polymer solutions.
It can be used for pollution degree detection, includes hydraulic oil, aviation kerosene, lubricant oil, transformer oil( insulation oil ), steam turbine oil, gear oil, engine oil, and water-based hydraulic oil etc.
| Parameter | Specification |
|---|---|
| Light source | Semiconducting laser narrow light detector (more accurate, steady and rapid) |
| Particle size range | 1μm ~ 400μm |
| Sensitivity | 1μm (ISO4402) or 4μm(C) (ISO11171) |
| Resolution | 10% |
| Off-line viscosity detection | ≤160cSt (With a pressure device can up to 400cSt) |
| Detection channel | ISO 11171: 4μm(C) ~ 70μm(C) 8 channels: 4, 6, 10, 14, 21, 25, 38, 70μm(C)ISO 4402: 1μm ~ 400μm 8 channels: 2, 5, 10, 15, 20, 25, 50, 100μm |
| Sampling volume | 10ml, sampling volume accuracy: >3% |
| Detection velocity | 5 ~ 80mL/min |
| Washing velocity | 5 ~ 80mL/min |
| Washing volume | 0ml ~ 90ml (minimum unit 1ml) |
| On-line detection pressure | 0.1 ~ 0.6MPa (With a pressure device can up to 41MPa) |
| Power supply | 24V Rechargeable lithium battery |
| Communication Interface | RS232 or RS485 |
| Testing data output | Equipped with a printer and supports USB download of historical data |
| Dimension | 435 × 304 × 170 mm |
| Net weight | 5 kg |
| Work environment temperature | -20°C ~ 60°C |
| Storage temperature | -30°C ~ 80°C |