Mail Us: hello@wrtindia.in

WPC150 Portable Particle Counter Basic, ADV & PRO Models

WPC150 Portable Particle Counter is made in the principle of light extinction (or light blockage) and is used to detect the size and quantity of solid particles in liquid. It is widely used in aviation, space, electricity, oil, chemical, transportation, port, metallurgy, machinery and automobile manufacturing for detecting contamination of solid particles for hydraulic oil, lubricating oil, transformer oil, turbo generator oil (turbine oil), gear oil, engine oil, jet fuel and water based hydraulic oil and insoluble particles for organic liquid and polymer solutions.

Applications

It can be used for pollution degree detection, includes hydraulic oil, aviation kerosene, lubricant oil, transformer oil( insulation oil ), steam turbine oil, gear oil, engine oil, and water-based hydraulic oil etc.

Product features

Technology advantage

Specification

Parameter Specification
Light source Semiconducting laser narrow light detector (more accurate, steady and rapid)
Particle size range 1μm ~ 400μm
Sensitivity 1μm (ISO4402) or 4μm(C) (ISO11171)
Resolution 10%
Off-line viscosity detection ≤160cSt (With a pressure device can up to 400cSt)
Detection channel ISO 11171: 4μm(C) ~ 70μm(C)
8 channels: 4, 6, 10, 14, 21, 25, 38, 70μm(C)ISO 4402: 1μm ~ 400μm
8 channels: 2, 5, 10, 15, 20, 25, 50, 100μm
Sampling volume 10ml, sampling volume accuracy: >3%
Detection velocity 5 ~ 80mL/min
Washing velocity 5 ~ 80mL/min
Washing volume 0ml ~ 90ml (minimum unit 1ml)
On-line detection pressure 0.1 ~ 0.6MPa (With a pressure device can up to 41MPa)
Power supply 24V Rechargeable lithium battery
Communication Interface RS232 or RS485
Testing data output Equipped with a printer and supports USB download of historical data
Dimension 435 × 304 × 170 mm
Net weight 5 kg
Work environment temperature -20°C ~ 60°C
Storage temperature -30°C ~ 80°C